The control code is broadcast commonly to all connected multiplexers at the same time. Hence, TEMES only knows one multiplexer (and treats the other multiplexers the same way). The settings are made in the path \Devices\<device> in the tab Multiplexer. The first checkbox must be checked to inform TEMES that a control byte has to be broadcast to connected multiplexers. The second checkbox must be checked if the multiplexer(s) are synchronized with the slow recording cycle instead of the fast recording cycle. A name can be given to the channel counter edit box which allows to refer to it inside TEMES. Then, the number of time slots must be entered (from 1 to 16, and 16 being the default value). The table consists of three columns: The first column is the time slot, the second column is the channel (from 1 to 16), and the third column is the operating range (from 1 to 16).
Operating Range |
Min. Temperature [°C] |
Max. Temperature [°C] |
---|---|---|
1 |
66.7 |
1372.0 |
2 |
50.3 |
1200.0 |
3 |
40.4 |
943.7 |
4 |
20.4 |
474.0 |
5 |
10.2 |
240.5 |
6 |
6.8 |
159.3 |
7 |
5.1 |
119.1 |
8 |
4.1 |
95.4 |
9 |
17.0 |
1372.0 |
10 |
0.0 |
1144.9 |
11 |
-10.4 |
892.6 |
12 |
-31.7 |
426.0 |
13 |
-42.8 |
189.0 |
14 |
-46.5 |
109.0 |
15 |
-48.4 |
70.0 |
16 |
-49.5 |
46.3 |
With linearized temperatures.