A voltage input must be chosen at which the thermocouple amplifier is connected. This is done in the path \Devices\<device>\Analog Input\Ax, and in the tab Multiplexer: It is recommended to check all three checkboxes and set the settling time to 20 ms (Note that the channel sample rate must be slower than 20 ms to obtain at least one value). A plus symbol appears in the parameter tree after pressing the ENTER key. By clicking this plus symbol further inputs are available which will be explained in the next section. However, the multiplexed signal needs a parameter definition for the main temperatures: This is done by selecting the tab Char. Curve and by choosing the table th2 N=29 (or th16mi N=31). The nonlinearized parameters are specified in the tab Settings, where first a name must be entered and an assignment according to the main temperature range must be made (see next section).