Calibrating a Temperature Multiplexer

Scroll Prev Top Next More

For convenience reasons, a separate parameter set is created with one multiplexer time slot which is set to the operating range to be calibrated. The sample rate is set to 200 ms (to obtain a very smooth signal averaging). Then, the multiplexed voltage input (and its first sub-channel) must be defined as described as before. And a display output for the multiplexed voltage input should be defined as well. Then, call menu item Hardware Calibrate device... and select the nonlinearized temperature signal, attach the voltage for the linearized temperature (e. g. with a thermocouple calibrator) and press the corresponding button of the two-point calibration. Attach the other linearized temperature and press the corresponding button of the two-point calibration. Then press OK. The device is now calibrated.

 

Operating Range

Tc1,n [°C]

Tc1 [°C]

Tc2,n [°C]

Tc2 [°C]

1

70.0418

70

1346.9403

1370

2

59.8673

60

1200

1200

3

49.6928

50

956.3775

940

4

29.5650

30

475.8663

470

5

19.6117

20

239.4937

240

6

9.7567

10

150.8233

150

7

9.7567

10

110.7889

110

8

9.7567

10

90.4645

90

9

19.6117

20

1346.9403

1370

10

0

0

1145.5394

1140

11

-9.6338

-10

907.4711

890

12

-28.4099

-30

423.7159

420

13

-37.5276

-40

180.3391

180

14

-37.5276

-40

100.6390

100

15

-37.5276

-40

59.8673

60

16

-37.5276

-40

39.5920

40

 

With first nonlinearized temperature Tc1,n, first linearized temperature Tc1, second nonlinearized temperature Tc2,n and second linearized temperature Tc2.