Requirements
calib-1.0.6.exe requires the Microsoft .NET Framework 4 which is part
of Windows 8 or newer, or which can be downloaded from
https://www.microsoft.com/en-us/download/details.aspx?id=17718. And the Microsoft .NET Framework 4 Full Language Pack can be downloaded from
https://www.microsoft.com/de-de/download/details.aspx?id=3324
.
Version history
V1.0.6
Support for signal converter USBDL1 GA.
V1.0.5
Support for signal converter SICO3M.
V1.0.4
Support for data logger USBDL1.
V1.0.3
Bugfix for version string reading.
V1.0.2
Optional saving of the calibrated data onto the disk (when the registry value "CfgPath" is defined with the storage path).
V1.0.1
User interface improvements
V1.0.0
First release
Usage Notes
Connect the device with the PC via a (serial) COM port.
Choose a (serial) COM port (Main menu → Settings → COM port).
Select your common inputs (= inputs with the same voltage level).
Press the continue button on the first tab.
Attach the corresponding reference voltage to the displayed inputs.
Press the input button which will take about 5-10 seconds to read in the corresponding voltages.
Repeat the process for the remaining reference voltage/inputs.
Optionally, select the output tab. Select the corresponding reference voltage and press the output button. Read the output voltage via a voltmeter and insert the reading into the table.
Choose the program tab and press the program device button.
The device is now newly calibrated. Please select menu item File → New in order to calibrate another device.